Vehicle Electrical Isolation Procedure
- 1076921-00-B Insulation Multimeter, Fluke 1507
- 1130480-00-A Test Probes, Slim, Fluke TP38
SPECIAL TOOLS
Insulation Multimeter, Fluke 1507 (1076921-00-B) |
Test Probes, Slim, Fluke TP38 (1130480-00-A) |
Only technicians who have been trained in High Voltage Awareness are permitted to perform this procedure. Proper personal protective equipment (PPE) and insulating HV gloves with a minimum rating of class 0 (1000V) must be worn at all times a high voltage cable, busbar, or fitting is handled. Refer to Tech Note TN-15-92-003, "High Voltage Awareness Care Points" for additional safety information.
Procedure
| 1 | Disconnect 12V power. See 12V Power (Disconnect and Connect). | ||||||||||||||||||||||||
| 2 | Set the insulation multimeter to measure the voltage across the 12V auxiliary battery terminals. Caution: If the voltage across the 12V auxiliary battery terminals is less than 10 volts or greater than 14 volts, the multimeter is not measuring reliably and must not be used. Use only a fully functional insulation multimeter.
| ||||||||||||||||||||||||
| 3 | Remove all jewelry and empty your pockets before continuing this procedure. | ||||||||||||||||||||||||
| 4 | Put on HV insulating gloves and leather outer gloves before continuing this procedure. | ||||||||||||||||||||||||
| 5 | Remove the HV probe cover. See Cover - HV Probe - HV Battery (Remove and Replace). | ||||||||||||||||||||||||
| 6 | Use the insulation multimeter to probe the voltages.
| ||||||||||||||||||||||||
| 7 | Install the HV probe cover. See Cover - HV Probe - HV Battery (Remove and Replace). | ||||||||||||||||||||||||
| 8 | Return to the procedure that required the vehicle to be electrically isolated. Note: After finishing the procedure that required the vehicle to be electrically isolated, perform a penthouse air leak test. See Penthouse Air Leak Test.
|